User Tools

Site Tools


howto:antibunching_measurements

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
howto:antibunching_measurements [2023/09/07 11:50] – [Changing from Normal Acquisition (T3-mode) to Antibunching (T2-mode)] peterhowto:antibunching_measurements [2023/09/07 11:54] (current) – [Changing from Normal Acquisition (T3-mode) to Antibunching (T2-mode)] peter
Line 29: Line 29:
 ==== Changing from Normal Acquisition (T3-mode) to Antibunching (T2-mode) ==== ==== Changing from Normal Acquisition (T3-mode) to Antibunching (T2-mode) ====
      
-=== Cabling: PicoHarp ===+=== Cabling: PicoHarp 300 ===
  
      -Disconnect Sync from [[Products:PicoHarp 300]] Ch.0.      -Disconnect Sync from [[Products:PicoHarp 300]] Ch.0.
Line 36: Line 36:
      -Adjustment of the CFD Level for Ch 0 (Sync Channel) might be necessary.       -Adjustment of the CFD Level for Ch 0 (Sync Channel) might be necessary. 
  
-    *It is also possible to connect both Detectors directly to the [[Products:PicoHarp 300]] without passing SPAD 2 through the router.    **In this case, however, the router cable (DVI-cable) needs to be disconnected from the PH300** +It is also possible to connect both detectors directly to the [[Products:PicoHarp 300]] without passing SPAD 2 through the router. **In this case, however, the router cable (DVI-cable) needs to be disconnected from PicoHarp300** 
-    *No special delay cables necessary +  
  
-=== Cabling: HydraHarp ===+=== Cabling: HydraHarp 400 ===
  
-    *Disconnect Sync from [[Products:HydraHarp 400]] Ch0. (thats all)+    *Disconnect the Sync signal from [[Products:HydraHarp 400]] SYNC IN. (That is all.)
  
-    *if Laser is attached via [[Products:SEPIA II]] driver, it's enough to disable the SYNC in the SEPIA software => no cable change necessary +    *if the laser is an LDH-series heasd driven by [[Products:SEPIA II]] driver, it's enough to disable the SYNC in the SEPIA software => no cable change necessary.
  
 === Software Settings during Acquisition === === Software Settings during Acquisition ===
howto/antibunching_measurements.txt · Last modified: 2023/09/07 11:54 by peter