howto:using_the_flcs_script_for_spectral_crosstalk_removal_via_flccs

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
howto:using_the_flcs_script_for_spectral_crosstalk_removal_via_flccs [2014/05/23 07:47] adminhowto:using_the_flcs_script_for_spectral_crosstalk_removal_via_flccs [2015/02/25 14:00] (current) bornemann
Line 1: Line 1:
 {{tag> howto tutorial FCS FLCS crosstalk analysis SymPhoTime}} {{tag> howto tutorial FCS FLCS crosstalk analysis SymPhoTime}}
-~~TOC~~+~~NOTOC~~
  
-====== Spectral crosstalk removal via FLCCS ======+====== Spectral Crosstalk Removal via FLCCS ======
  
  
Line 9: Line 9:
  
 This tutorial shows step-by-step, how to calculate a spectral fluorescence lifetime cross-correlation (FLCCS) curve from a measurement of a mixture of two labeled oligunucleotides moving independently from each other. As the green dye has been excited with pulsed 485 nm excitation, while the red dye has been excited with a 560 nm cw laser, the different excitation scheme is used to eliminate the unwanted spectral crosstalk. This tutorial shows step-by-step, how to calculate a spectral fluorescence lifetime cross-correlation (FLCCS) curve from a measurement of a mixture of two labeled oligunucleotides moving independently from each other. As the green dye has been excited with pulsed 485 nm excitation, while the red dye has been excited with a 560 nm cw laser, the different excitation scheme is used to eliminate the unwanted spectral crosstalk.
 +
 +{{ youtube>O2lTTF70tZE?large }}
  
 ===== Background Information ===== ===== Background Information =====
Line 134: Line 136:
 **Response:** The correlation curves of this file are calculated. The cross-correlation amplitude is significantly higher than for the negative calibration dye as processed before, but lower than the corresponding FCCS file obtained with the "Grouped FCS" script. **Response:** The correlation curves of this file are calculated. The cross-correlation amplitude is significantly higher than for the negative calibration dye as processed before, but lower than the corresponding FCCS file obtained with the "Grouped FCS" script.
 {{ using_the_flcs_script_for_spectral_crosstalk_removal_via_flccs_Image_18.png?600 }} {{ using_the_flcs_script_for_spectral_crosstalk_removal_via_flccs_Image_18.png?600 }}
- 
  
howto/using_the_flcs_script_for_spectral_crosstalk_removal_via_flccs.txt · Last modified: 2015/02/25 14:00 by bornemann