howto:t3r_antibunching_-_slow_decay [2022/02/01 10:52] – [Antibunching by histogramming - Slow Decay] peterhowto:t3r_antibunching_-_slow_decay [2022/02/01 10:53] – [In this show-case the external preconditions are:] peter
Line 3: Line 3:
 ====== Artefacts in Antibunching Histograms - Slow Decay ====== ====== Artefacts in Antibunching Histograms - Slow Decay ======
  
-===== Q.: How can we avoid decay at very long time delay (up to 5us)? ===== 
-==== In this show-case the external preconditions are: ==== 
  
-- detectors: perkin elmer SPCMAQR14 
- 
-- 700 nm SP filter in front of one of the APDs 
- 
-- sample: diamond nanocrystals (100 nm), we are measuring the antibunching of the NV centers emission 
- 
-- laser wavelength: 532 nm, cw 
- 
-- T3R Measurement with PicoHarp Software  
  
 ===== A.: The decay can be caused by blinking of the nanodiamonds. ===== ===== A.: The decay can be caused by blinking of the nanodiamonds. =====