howto:how_to_measure_the_instrument_response_function_irf
Differences
This shows you the differences between two versions of the page.
Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
howto:how_to_measure_the_instrument_response_function_irf [2016/12/06 11:50] – veiga | howto:how_to_measure_the_instrument_response_function_irf [2023/09/07 22:55] (current) – [Using samples with ultrafast decay] peter | ||
---|---|---|---|
Line 4: | Line 4: | ||
====== How to Measure the Instrument Response Function (IRF) ====== | ====== How to Measure the Instrument Response Function (IRF) ====== | ||
- | (For an explanation of the term IRF see [[glossary: | + | (For an explanation of the term Instrument Response Function |
+ | |||
+ | The following video shows how to measure na IRF on confocal microscopes. | ||
{{youtube> | {{youtube> | ||
+ | |||
+ | \\ | ||
- | ==== Make sure that the detection | + | ==== Low count rate during IRF measurements |
- | Diluting the scattering solution is better than using grey filters. Ideal is when the decay and the IRF are recorded at the same [[glossary: | + | Make sure that the detection count rate is much lower than the count rate during a fluorescence decay measurement. |
- | If the IRF should be measured on a microscope system with SPAD detectors, | + | If the IRF should be measured |
===== Using samples with ultrafast decay ===== | ===== Using samples with ultrafast decay ===== | ||
- | Some detectors (particularly SPADs) have wavelength dependent timing response. In this case an IRF recorded at the excitation wavelength may not be useful for precise reconvolution. The solution is to acquire the IRF at the fluorescence wavelength, or at least spectrally closer to the fluorescence emission. | + | Some detectors (particularly |
==== General recipe ==== | ==== General recipe ==== | ||
Line 43: | Line 47: | ||
| | ||
add 0.17 mL of saturated water solution of Erythrosine B (at least 95% of purity) | add 0.17 mL of saturated water solution of Erythrosine B (at least 95% of purity) | ||
- | add 0.03 mL of 0.004 M KOH (potassium hydroxide) solution in order to achieve pH10 | ||
</ | </ | ||
Line 64: | Line 67: | ||
Put a droplet on a coverslip, measurement conditions as for fluorescence measurement | Put a droplet on a coverslip, measurement conditions as for fluorescence measurement | ||
</ | </ | ||
- | ((Szabelski M., Iliev D., Sarkar P., Luchowski R., Gryczynski Z., Kapusta P., Erdmann R., Gryczinski I.\\ | + | |
- | Collisional quenching | + | |
- | Applied Spectroscopy, Vol.63, p.0363-0368 (2009)\\ | + | ==== Two photon excitation |
- | http://www.ingentaconnect.com/ | + | |
+ | Do not attempt to record an [[glossary: | ||
+ | |||
+ | You can try to excite (by [[glossary: | ||
+ | |||
+ | With microscopes it is convenient to record the second harmonic signal that is generated on the surface of urea crystals. The best is to let evaporate a droplet of concentrated urea solution on a clear cover slip. The resulting film of micro-crystals is easy to target. | ||
+ | |||
+ | Urea, aka Carbamide or Carbonyldiamide, | ||
+ | |||
+ | ===== Appropriate Count Rate for Measuring an IRF ===== | ||
+ | See [[glossary: | ||
+ | |||
+ | ===== How often does the IRF need to be measured? ===== | ||
+ | |||
+ | In spectrometers, | ||
+ | |||
+ | In microscopy-applications, this is usually not practical; and often it is sufficient to measure the IRF once during a measurement series, provided that the system has had time to warm up (~15min), and neither the repetition rate nor the intensity at the diode laser driver is changed (the current which drives the laser). | ||
+ | If the intensity needs to be changed, the optical attenuation can be adapted. | ||
+ | |||
+ | A special case are systems with 2-Photon-Excitation (2PE). Here, usually TiSa-lasers are used which have fs-pulses, therefore the IRF is normally determined by the detector. In these cases, often the IRF can be measured once (and the excitation wavelength is not important, provided that the IRF is measured with a quenched dye and the same filterset is used as for the sample), and re-used later. Over time or upon changes of the excitation wavelength, the position of the IRF can shift slightly, but this is accounted for with the " | ||
+ | |||
+ | |||
+ | ===== How to compensate IRF effects in the analysis | ||
+ | |||
+ | There are two major ways of compensating IRF effects: | ||
+ | |||
+ | *correct the effects in the data (**de**convolution) | ||
+ | *take the effects into account in your model equation (**re**convolution) | ||
+ | |||
+ | Note: All analysis packages from PicoQuant use the [[glossary: | ||
+ | |||
+ | ===== Measuring the IRF as scattered excitation light ===== | ||
+ | |||
+ | We do not recommend to measure the IRF as scatters light in microscopy, due to the color dependence of SPAD detectors, which are generally used in microscopy. Furthermore, | ||
+ | |||
+ | However, in case of cuvette based measurement like in spectrometers, | ||
+ | |||
+ | Note that recording the IRF via scattering requires tuning the emission monochromator to the excitation wavelength. In filter based machines, e.g. FluoTime100 this means removing the emission bandpass or longpass filter. In microscopes, | ||
- | ==== Selected literature | + | ==== Selected literature: ==== |
Luchowski R., Kapusta P., Szabelski M., Sarkar P., Borejdo J., Gryczynski Z., Gryczynski I.\\ | Luchowski R., Kapusta P., Szabelski M., Sarkar P., Borejdo J., Gryczynski Z., Gryczynski I.\\ | ||
Line 86: | Line 126: | ||
Evaluation of instrument response functions for lifetime imaging detectors using quenched Rose Bengal solutions\\ | Evaluation of instrument response functions for lifetime imaging detectors using quenched Rose Bengal solutions\\ | ||
Chemical Physics Letters, Vol.471, p.153-159 (2009)\\ | Chemical Physics Letters, Vol.471, p.153-159 (2009)\\ | ||
- | http://dx.doi.org/10.1016/j.cplett.2009.02.001 | + | https://www.sciencedirect.com/science/article/ |
Line 92: | Line 132: | ||
Collisional quenching of Erythrosine B as a potential reference dye for impulse response function evaluation\\ | Collisional quenching of Erythrosine B as a potential reference dye for impulse response function evaluation\\ | ||
Applied Spectroscopy, | Applied Spectroscopy, | ||
- | http://www.ingentaconnect.com/content/sas/ | + | https://www.osapublishing.org/as/viewmedia.cfm? |
Line 98: | Line 138: | ||
Photophysical properties of novel fluorescein derivative and its applications for time-resolved fluorescence spectroscopy\\ | Photophysical properties of novel fluorescein derivative and its applications for time-resolved fluorescence spectroscopy\\ | ||
Chemical Physics Letters, Vol.493, p.399-403 (2010)\\ | Chemical Physics Letters, Vol.493, p.399-403 (2010)\\ | ||
- | http://dx.doi.org/10.1016/j.cplett.2010.05.061 | + | https://www.sciencedirect.com/science/article/ |
Line 108: | Line 148: | ||
Picosecond fluorescence of intact and dissolved PSI-LHCI crystals\\ | Picosecond fluorescence of intact and dissolved PSI-LHCI crystals\\ | ||
Biophysical Journal, Vol.95, p.5851-5861 (2008)\\ | Biophysical Journal, Vol.95, p.5851-5861 (2008)\\ | ||
- | http://dx.doi.org/10.1529/biophysj.108.140467 | + | https://www.sciencedirect.com/science/article/ |
Line 135: | Line 175: | ||
- | ==== Two photon excitation (TPE) ==== | ||
- | |||
- | Do not attempt to record an [[glossary: | ||
- | |||
- | You can try to excite (by [[glossary: | ||
- | |||
- | With microscopes it is convenient to record the second harmonic signal that is generated on the surface of urea crystals. The best is to let evaporate a droplet of concentrated urea solution on a clear cover slip. The resulting film of micro-crystals is easy to target. | ||
- | |||
- | Urea, aka Carbamide or Carbonyldiamide, | ||
- | |||
- | ===== Appropriate Count Rate for Measuring an IRF ===== | ||
- | See [[glossary: | ||
- | |||
- | ===== How to compensate IRF effects in the analysis of time domain measurements ===== | ||
- | |||
- | There are two major ways of compensating IRF effects: | ||
- | |||
- | *correct the effects in the data (**de**convolution) | ||
- | *take the effects into account in your model equation (**re**convolution) | ||
- | |||
- | Note: All analysis packages from PicoQuant use the [[glossary: | ||
- | |||
- | ===== Measuring the IRF as scattered excitation light ===== | ||
- | |||
- | We do not recommend to measure the IRF as scatters light in microscopy, due to the color dependence of SPAD detectors, which are generally used in microscopy. Furthermore, | ||
- | |||
- | However, in case of cuvette based measurement like in spectrometers, | ||
- | |||
- | Note that recording the IRF via scattering requires tuning the emission monochromator to the excitation wavelength. In filter based machines, e.g. FluoTime100 this means removing the emission bandpass or longpass filter. In microscopes, | ||
howto/how_to_measure_the_instrument_response_function_irf.1481025027.txt.gz · Last modified: 2016/12/06 11:50 by veiga