howto:how_to_measure_the_instrument_response_function_irf [2021/08/06 07:57] buschmannhowto:how_to_measure_the_instrument_response_function_irf [2022/01/24 10:22] – [IRF measurement with KI and Erythrosine B] rezvani
Line 43: Line 43:
  prepare 1mL of saturated water solution of KI (potassium iodide)   prepare 1mL of saturated water solution of KI (potassium iodide) 
  add 0.17 mL of saturated water solution of Erythrosine B (at least 95% of purity)  add 0.17 mL of saturated water solution of Erythrosine B (at least 95% of purity)
- add 0.03 mL of 0.004 M KOH (potassium hydroxide) solution in order to achieve pH10  
 </code> </code>
  
Line 145: Line 144:
 Picosecond fluorescence of intact and dissolved PSI-LHCI crystals\\  Picosecond fluorescence of intact and dissolved PSI-LHCI crystals\\ 
 Biophysical Journal, Vol.95, p.5851-5861 (2008)\\  Biophysical Journal, Vol.95, p.5851-5861 (2008)\\ 
-http://dx.doi.org/10.1529/biophysj.108.140467+https://www.sciencedirect.com/science/article/pii/S0006349508820012
  
  