User Tools

Site Tools


howto:how_to_measure_the_instrument_response_function_irf

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
Next revisionBoth sides next revision
howto:how_to_measure_the_instrument_response_function_irf [2016/10/18 11:40] – [Selected literature:] buschmannhowto:how_to_measure_the_instrument_response_function_irf [2019/11/04 09:45] – [Appropriate Count Rate for Measuring an IRF] buschmann
Line 8: Line 8:
 {{youtube>cjSgCBe3ReM?large}} {{youtube>cjSgCBe3ReM?large}}
  
- 
-===== Using scattered excitation light ===== 
- 
- 
-In case of cuvette based measurement the simplest procedure is to use a very diluted solution of colloidal silica. (LUDOX is often used, LUDOX is a trade mark by DuPont. It can be purchased via Aldrich or Sigma.) Do not use "non-diary cafe creamer" or glycogen, mentioned in older literature. These compunds are fluorescent. The scattering solution must be really weak, typical starting "concentration" is one droplet of the colloid from the original LUDOX bottle (as delivered with you system) diluted 100x. If the signal is too strong, dilute further. 
- 
-Note that recording the IRF via scattering requires tuning the emission monochromator to the excitation wavelength. In filter based machines, e.g. FluoTime100 this means removing the emission bandpass or longpass filter. In microscopes, this is equivalent to replacing the detection filter with an OD filter (OD = optical density, grey filter) and recording the back scattering from e.g. a clean glass cover slip. Beware, on some confocal LSMs it is simply not possible to measure at the excitation wl. due to the restrictions introduced into the system software by the manufacturer. In case of measurements of solid samples in FT100 or FT200 spectrometers, the surface scattering is usually so strong that it is not necessary to mount a special sample for IRF. Tuning the mono to the exc. wl. and attenuating the excitation beam is sufficient.  
  
 ==== Make sure that the detection count rate is much lower than the count rate used for fluorescence decay measurement. ==== ==== Make sure that the detection count rate is much lower than the count rate used for fluorescence decay measurement. ====
Line 26: Line 19:
  
 Some detectors (particularly SPADs) have wavelength dependent timing response. In this case an IRF recorded at the excitation wavelength may not be useful for precise reconvolution. The solution is to acquire the IRF at the fluorescence wavelength, or at least spectrally closer to the fluorescence emission.  Some detectors (particularly SPADs) have wavelength dependent timing response. In this case an IRF recorded at the excitation wavelength may not be useful for precise reconvolution. The solution is to acquire the IRF at the fluorescence wavelength, or at least spectrally closer to the fluorescence emission. 
 +
 +==== General recipe ====
 +
 +  * Create a saturated aqueous KI (potassium iodide) solution. (Note, NOT KCl, but iodide.) Beware that KI is pretty well soluble in water. You will need quite a lot of KI. Our experience is that the volume of KI crystals is almost the same as the volume of the water added. The solution MUST be saturated.
 +
 +  * Then add any water soluble fluorescent dye with emission spectrum in the range where you need to record the IRF.
 +
 +  * Good luck with IRF measurements. Do not expect high count rates, but IRF must be recorded at much lower (than fluorescence) count rates anyway.
 +
 +Our guess is that this quenching trick works generally. Specifically, we found that classic Fluorescein works at 480-500 emission with 440 nm excitation. Alexas or ATTOs would be perhaps even better, although more expensive...
 +
 +Note: the strongly quenched solution will not look fluorescent when watched by eye. However, it must have a strong color (strong absorbtion, high dye concentration). Nevertheless, be careful when adding the dye, it must be completely dissolved.
 +
 +Such a cocktail cannot be stored for a long time. Latest the KI will photochemically decompose.
 +
 +==== IRF measurement with KI and Erythrosine B ====
 +
 +SPAD detectors have wavelengths dependent timing response. IRF recorded at the excitation wavelength may not be useful for precise re-convolution. Using Erythrosine B, the IRF is acquired at the fluorescence wavelength.
 +
 +Recipe:
 +
 +<code>
 + prepare 1mL of saturated water solution of KI (potassium iodide) 
 + add 0.17 mL of saturated water solution of Erythrosine B (at least 95% of purity)
 + add 0.03 mL of 0.004 M KOH (potassium hydroxide) solution in order to achieve pH10 
 +</code>
 +
 +Storage:
 +
 +<code>
 +keep at ~ 4°C, renew the solution after one month
 +</code>
 +
 +Spectra:
 +
 +<code>
 +excitation from 470 nm to 540 nm
 +emission from 500 nm to 600 nm
 +</code>
 +
 +Measurement:
 +
 +<code>
 +Put a droplet on a coverslip, measurement conditions as for fluorescence measurement 
 +</code>
 + 
 +
 +==== Two photon excitation (TPE) ====
 +
 +Do not attempt to record an [[glossary:IRF]] at the fundamental (IR) wavelength. The resulting pulse form would be meaningless anyway.
 +
 +You can try to excite (by [[glossary:TPE]], of course) any of the above mentioned fast decay time fluorophores and record their response. The signal will be wery weak, but this is not a problem for IRF measurements.
 +
 +With microscopes it is convenient to record the second harmonic signal that is generated on the surface of urea crystals. The best is to let evaporate a droplet of concentrated urea solution on a clear cover slip. The resulting film of micro-crystals is easy to target.
 +
 +Urea, aka Carbamide or Carbonyldiamide, CAS Number:57-13-6
 +
 +===== Appropriate Count Rate for Measuring an IRF =====
 +See [[glossary:Differential Count Rate]]
 +
 +===== How often does the IRF need to be measured? =====
 +
 +In spectrometers, the IRF is usually measured before or after each measurement.
 +
 +In microscopy-applications, this is usually not practical; and often it is sufficient to measure the IRF once during a measurement series, provided that the system has had time to warm up (~15min), and neither the repetition rate nor the intensity at the diode laser driver is changed (the current which drives the laser). 
 +If the intensity needs to be changed, the optical attenuation can be adapted.
 +
 +A special case are systems with 2-Photon-Excitation (2PE). Here, usually TiSa-lasers are used which have fs-pulses, therefore the IRF is normally determined by the detector. In these cases, often the IRF can be measured once (and the excitation wavelength is not important, provided that the IRF is measured with a quenched dye and the same filterset is used as for the sample), and re-used later. Over time or upon changes of the excitation wavelength, the position of the IRF can shift slightly, but this is accounted for with the "IRF-shift"-parameter of the fitting equation. 
 +
 +
 +===== How to compensate IRF effects in the analysis of time domain measurements =====
 +
 +There are two major ways of compensating IRF effects:
 +
 +    *correct the effects in the data (**de**convolution)
 +    *take the effects into account in your model equation (**re**convolution) 
 +
 +Note: All analysis packages from PicoQuant use the [[glossary:reconvolution]] method. The reason is simple: [[glossary:Deconvolution]] has many disadvantages: It is rather slow, it is an ill posed problem producing a lot of noise (at best), it does not get you any nearer to the model parameters, and it is very likely to produce artefacts without allowing to tell apart artefacts and effects.
 +
 +===== Measuring the IRF as scattered excitation light =====
 +
 +We do not recommend to measure the IRF as scatters light in microscopy, due to the color dependence of SPAD detectors, which are generally used in microscopy. Furthermore, even with detectors lacking color effects, it is disadvantageous, due to the many reflection peaks found along the light pathway in a microscope set up. 
 +
 +However, in case of cuvette based measurement like in spectrometers, this is  the method of choice. The simplest procedure is to use a very diluted solution of colloidal silica. (LUDOX is often used, LUDOX is a trade mark by DuPont. It can be purchased via Aldrich or Sigma.) Do not use "non-diary cafe creamer" or glycogen, mentioned in older literature. These compunds are fluorescent. The scattering solution must be really weak, typical starting "concentration" is one droplet of the colloid from the original LUDOX bottle (as delivered with you system) diluted 100x. If the signal is too strong, dilute further.
 +
 +Note that recording the IRF via scattering requires tuning the emission monochromator to the excitation wavelength. In filter based machines, e.g. FluoTime100 this means removing the emission bandpass or longpass filter. In microscopes, this is equivalent to replacing the detection filter with an OD filter (OD = optical density, grey filter) and recording the back scattering from e.g. a clean glass cover slip. Beware, on some confocal LSMs it is simply not possible to measure at the excitation wl. due to the restrictions introduced into the system software by the manufacturer. In case of measurements of solid samples in FT100 or FT200 spectrometers, the surface scattering is usually so strong that it is not necessary to mount a special sample for IRF. Tuning the mono to the exc. wl. and attenuating the excitation beam is sufficient. 
  
 ==== Selected literature: ==== ==== Selected literature: ====
Line 91: Line 170:
 Applied Spectroscopy, Vol. 68, p. 577-583 (2014) Applied Spectroscopy, Vol. 68, p. 577-583 (2014)
 http://dx.doi.org/10.1366/13-07236 http://dx.doi.org/10.1366/13-07236
- 
- 
    
  
-==== General recipe ==== 
  
-  * Create a saturated aqueous KI (potassium iodide) solution. (Note, NOT KCl, but iodide.) Beware that KI is pretty well soluble in water. You will need quite a lot of KI. Our experience is that the volume of KI crystals is almost the same as the volume of the water added. The solution MUST be saturated. 
  
-  * Then add any water soluble fluorescent dye with emission spectrum in the range where you need to record the IRF. 
- 
-  * Good luck with IRF measurements. Do not expect high count rates, but IRF must be recorded at much lower (than fluorescence) count rates anyway. 
- 
-Our guess is that this quenching trick works generally. Specifically, we found that classic Fluorescein works at 480-500 emission with 440 nm excitation. Alexas or ATTOs would be perhaps even better, although more expensive... 
- 
-Note: the strongly quenched solution will not look fluorescent when watched by eye. However, it must have a strong color (strong absorbtion, high dye concentration). Nevertheless, be careful when adding the dye, it must be completely dissolved. 
- 
-Such a cocktail cannot be stored for a long time. Latest the KI will photochemically decompose. 
- 
-==== IRF measurement with KI and Erythrosine B ==== 
- 
-SPAD detectors have wavelengths dependent timing response. IRF recorded at the excitation wavelength may not be useful for precise re-convolution. Using Erythrosine B, the IRF is acquired at the fluorescence wavelength. 
- 
-Recipe: 
- 
-<code> 
- prepare 1mL of saturated water solution of KI (potassium iodide)  
- add 0.17 mL of saturated water solution of Erythrosine B (at least 95% of purity) 
- add 0.03 mL of 0.004 M KOH (potassium hydroxide) solution in order to achieve pH10  
-</code> 
- 
-Storage: 
- 
-<code> 
-keep at ~ 4°C, renew the solution after one month 
-</code> 
- 
-Spectra: 
- 
-<code> 
-excitation from 470 nm to 540 nm 
-emission from 500 nm to 600 nm 
-</code> 
- 
-Measurement: 
- 
-<code> 
-Put a droplet on a coverslip, measurement conditions as for fluorescence measurement  
-</code> 
- ((Szabelski M., Iliev D., Sarkar P., Luchowski R., Gryczynski Z., Kapusta P., Erdmann R., Gryczinski I.\\  
-Collisional quenching of Erythrosine B as a potential reference dye for impulse response function evaluation\\  
-Applied Spectroscopy, Vol.63, p.0363-0368 (2009)\\  
-http://www.ingentaconnect.com/content/sas/sas/2009/00000063/00000003/art00017))  
- 
-==== Two photon excitation (TPE) ==== 
- 
-Do not attempt to record an [[glossary:IRF]] at the fundamental (IR) wavelength. The resulting pulse form would be meaningless anyway. 
- 
-You can try to excite (by [[glossary:TPE]], of course) any of the above mentioned fast decay time fluorophores and record their response. The signal will be wery weak, but this is not a problem for IRF measurements. 
- 
-With microscopes it is convenient to record the second harmonic signal that is generated on the surface of urea crystals. The best is to let evaporate a droplet of concentrated urea solution on a clear cover slip. The resulting film of micro-crystals is easy to target. 
- 
-Urea, aka Carbamide or Carbonyldiamide, CAS Number:57-13-6 
- 
-===== Appropriate Count Rate for Measuring an IRF ===== 
-See [[glossary:Differential Count Rate]] 
-  
-===== How to compensate IRF effects in the analysis of time domain measurements ===== 
- 
-There are two major ways of compensating IRF effects: 
- 
-    *correct the effects in the data (**de**convolution) 
-    *take the effects into account in your model equation (**re**convolution)  
- 
-Note: All analysis packages from PicoQuant use the [[glossary:reconvolution]] method. The reason is simple: [[glossary:Deconvolution]] has many disadvantages: It is rather slow, it is an ill posed problem producing a lot of noise (at best), it does not get you any nearer to the model parameters, and it is very likely to produce artefacts without allowing to tell apart artefacts and effects. 
howto/how_to_measure_the_instrument_response_function_irf.txt · Last modified: 2023/09/07 22:55 by peter