User Tools

Site Tools


howto:how_to_measure_the_instrument_response_function_irf

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
Next revisionBoth sides next revision
howto:how_to_measure_the_instrument_response_function_irf [2016/01/04 13:46] – New dye, Allura Red added peterhowto:how_to_measure_the_instrument_response_function_irf [2019/11/04 09:45] – [Appropriate Count Rate for Measuring an IRF] buschmann
Line 8: Line 8:
 {{youtube>cjSgCBe3ReM?large}} {{youtube>cjSgCBe3ReM?large}}
  
- 
-===== Using scattered excitation light ===== 
- 
- 
-In case of cuvette based measurement the simplest procedure is to use a very diluted solution of colloidal silica. (LUDOX is often used, LUDOX is a trade mark by DuPont. It can be purchased via Aldrich or Sigma.) Do not use "non-diary cafe creamer" or glycogen, mentioned in older literature. These compunds are fluorescent. The scattering solution must be really weak, typical starting "concentration" is one droplet of the colloid from the original LUDOX bottle (as delivered with you system) diluted 100x. If the signal is too strong, dilute further. 
- 
-Note that recording the IRF via scattering requires tuning the emission monochromator to the excitation wavelength. In filter based machines, e.g. FluoTime100 this means removing the emission bandpass or longpass filter. In microscopes, this is equivalent to replacing the detection filter with an OD filter (OD = optical density, grey filter) and recording the back scattering from e.g. a clean glass cover slip. Beware, on some confocal LSMs it is simply not possible to measure at the excitation wl. due to the restrictions introduced into the system software by the manufacturer. In case of measurements of solid samples in FT100 or FT200 spectrometers, the surface scattering is usually so strong that it is not necessary to mount a special sample for IRF. Tuning the mono to the exc. wl. and attenuating the excitation beam is sufficient.  
  
 ==== Make sure that the detection count rate is much lower than the count rate used for fluorescence decay measurement. ==== ==== Make sure that the detection count rate is much lower than the count rate used for fluorescence decay measurement. ====
Line 26: Line 19:
  
 Some detectors (particularly SPADs) have wavelength dependent timing response. In this case an IRF recorded at the excitation wavelength may not be useful for precise reconvolution. The solution is to acquire the IRF at the fluorescence wavelength, or at least spectrally closer to the fluorescence emission.  Some detectors (particularly SPADs) have wavelength dependent timing response. In this case an IRF recorded at the excitation wavelength may not be useful for precise reconvolution. The solution is to acquire the IRF at the fluorescence wavelength, or at least spectrally closer to the fluorescence emission. 
- 
-==== Selected literature: ==== 
- 
-Luchowski R., Kapusta P., Szabelski M., Sarkar P., Borejdo J., Gryczynski Z., Gryczynski I.\\  
-Förster resonance energy transfer (FRET)-based picosecond lifetime reference for instrument response evaluation \\  
-Measurement Science and Technology, Vol.20, 095601 (2009)\\  
-http://dx.doi.org/10.1088/0957-0233/20/9/095601 
- 
- 
-Luchowski R., Gryczynski Z., Sarkar P., Borejdo J., Szabelski M., Kapusta P., Gryczynski I.\\  
-Instrument response standard in time-resolved fluorescence\\  
-Review of Scientific Instruments, Vol.80, 033109 (2009)\\  
-http://dx.doi.org/10.1063/1.3095677 
- 
- 
-Szabelski M., Luchowski R., Gryczynski Z., Kapusta P., Ortmann U., Gryczynski I.\\  
-Evaluation of instrument response functions for lifetime imaging detectors using quenched Rose Bengal solutions\\  
-Chemical Physics Letters, Vol.471, p.153-159 (2009)\\  
-http://dx.doi.org/10.1016/j.cplett.2009.02.001 
- 
- 
-Szabelski M., Iliev D., Sarkar P., Luchowski R., Gryczynski Z., Kapusta P., Erdmann R., Gryczinski I.\\  
-Collisional quenching of Erythrosine B as a potential reference dye for impulse response function evaluation\\  
-Applied Spectroscopy, Vol.63, p.0363-0368 (2009)\\  
-http://www.ingentaconnect.com/content/sas/sas/2009/00000063/00000003/art00017 
- 
- 
-Szabelski M., Gryczynski Z., Gryczynski I.\\  
-Photophysical properties of novel fluorescein derivative and its applications for time-resolved fluorescence spectroscopy\\  
-Chemical Physics Letters, Vol.493, p.399-403 (2010)\\  
-http://dx.doi.org/10.1016/j.cplett.2010.05.061 
- 
- 
-The dye **pinacyanol iodide** has a very fast decay time (in methanol 6 ps) and very broad emission spectrum (starting at around 550nm), thus suitable for determination of the IRF in a very broad range. 
- 
-Pinacyanol iodide is also known as Quinaldine blue, 1,1' -Diethyl-2,2' -carbocyanine iodide, CAS Number:605-91-4. Pinacyanol bromide, CAS Number:2670-67-9 or Pinacyanol chloride CAS Number:2768-90-3 should work too, however, the dye must be pure, otherwise you may end up with measuring (long lifetime) impurity emission. For details see: 
- 
-van Oort B., Amunts A., Borst J. W., van Hoek A., Nelson N., van Amerongen H., Croce R.\\  
-Picosecond fluorescence of intact and dissolved PSI-LHCI crystals\\  
-Biophysical Journal, Vol.95, p.5851-5861 (2008)\\  
-http://dx.doi.org/10.1529/biophysj.108.140467 
- 
- 
-The dye **Allura Red** has also a remarkably short fluorescence lifetime of about 10 ps. This short lifetime does not depend on the emission (observation) wavelength (from 550 nm to 750 nm) and it is another candidate for impulse response measurements. 
- 
-Allura Red is Disodium 6-hydroxy-5-[(2-methoxy-5-methyl-4-sulfophenyl)azo]-2-naphthalenesulfonate, CAS Number: 25956-17-6, Colour Index Number 16035, EC Number 247-368-0, MDL number MFCD00059526, PubChem Substance ID: 24869338\\ 
- 
-Rahul Chib, Sunil Shah, Zygmunt Gryczynski, Rafal Fudala, Julian Borejdo, Bogumil Zelent, Maria G Corradini, Richard D Ludescher, Ignacy Gryczynski\\  
-Standard reference for instrument response function in fluorescence lifetime measurements in visible and near infrared\\  
-Measurement Science and Technology, Vol.27, 027001 (2016)\\  
-http://dx.doi.org/10.1088/0957-0233/27/2/027001 
- 
- 
- 
-Even dye with lifetimes that are usually in the ns range can be used, if the lifetime can be quenched to a few ps. Quenching can be achieved using a saturated iodide solution. For details see: 
- 
-Mengwei Liu, Mengui Jia, Haifeng Pan, Lei Li, Mengfang Chang, Hua Ren, Françoise Argoul, Sanjun Zhang, Jianhua Xu\\ 
-Instrument Response Standard in Time-Resolved Fluorescence Spectroscopy at Visible Wavelength: Quenched Fluorescein Sodium\\ 
-Applied Spectroscopy, Vol. 68, p. 577-583 (2014) 
-http://dx.doi.org/10.1366/13-07236 
- 
- 
-  
  
 ==== General recipe ==== ==== General recipe ====
Line 134: Line 64:
 Put a droplet on a coverslip, measurement conditions as for fluorescence measurement  Put a droplet on a coverslip, measurement conditions as for fluorescence measurement 
 </code> </code>
- ((Szabelski M., Iliev D., Sarkar P., Luchowski R., Gryczynski Z., Kapusta P., Erdmann R., Gryczinski I.\\  + 
-Collisional quenching of Erythrosine B as a potential reference dye for impulse response function evaluation\\  +
-Applied Spectroscopy, Vol.63, p.0363-0368 (2009)\\  +
-http://www.ingentaconnect.com/content/sas/sas/2009/00000063/00000003/art00017)) +
  
 ==== Two photon excitation (TPE) ==== ==== Two photon excitation (TPE) ====
Line 151: Line 78:
 ===== Appropriate Count Rate for Measuring an IRF ===== ===== Appropriate Count Rate for Measuring an IRF =====
 See [[glossary:Differential Count Rate]] See [[glossary:Differential Count Rate]]
- + 
 +===== How often does the IRF need to be measured? ===== 
 + 
 +In spectrometers, the IRF is usually measured before or after each measurement. 
 + 
 +In microscopy-applications, this is usually not practical; and often it is sufficient to measure the IRF once during a measurement series, provided that the system has had time to warm up (~15min), and neither the repetition rate nor the intensity at the diode laser driver is changed (the current which drives the laser).  
 +If the intensity needs to be changed, the optical attenuation can be adapted. 
 + 
 +A special case are systems with 2-Photon-Excitation (2PE). Here, usually TiSa-lasers are used which have fs-pulses, therefore the IRF is normally determined by the detector. In these cases, often the IRF can be measured once (and the excitation wavelength is not important, provided that the IRF is measured with a quenched dye and the same filterset is used as for the sample), and re-used later. Over time or upon changes of the excitation wavelength, the position of the IRF can shift slightly, but this is accounted for with the "IRF-shift"-parameter of the fitting equation.  
 + 
 ===== How to compensate IRF effects in the analysis of time domain measurements ===== ===== How to compensate IRF effects in the analysis of time domain measurements =====
  
Line 160: Line 97:
  
 Note: All analysis packages from PicoQuant use the [[glossary:reconvolution]] method. The reason is simple: [[glossary:Deconvolution]] has many disadvantages: It is rather slow, it is an ill posed problem producing a lot of noise (at best), it does not get you any nearer to the model parameters, and it is very likely to produce artefacts without allowing to tell apart artefacts and effects. Note: All analysis packages from PicoQuant use the [[glossary:reconvolution]] method. The reason is simple: [[glossary:Deconvolution]] has many disadvantages: It is rather slow, it is an ill posed problem producing a lot of noise (at best), it does not get you any nearer to the model parameters, and it is very likely to produce artefacts without allowing to tell apart artefacts and effects.
 +
 +===== Measuring the IRF as scattered excitation light =====
 +
 +We do not recommend to measure the IRF as scatters light in microscopy, due to the color dependence of SPAD detectors, which are generally used in microscopy. Furthermore, even with detectors lacking color effects, it is disadvantageous, due to the many reflection peaks found along the light pathway in a microscope set up. 
 +
 +However, in case of cuvette based measurement like in spectrometers, this is  the method of choice. The simplest procedure is to use a very diluted solution of colloidal silica. (LUDOX is often used, LUDOX is a trade mark by DuPont. It can be purchased via Aldrich or Sigma.) Do not use "non-diary cafe creamer" or glycogen, mentioned in older literature. These compunds are fluorescent. The scattering solution must be really weak, typical starting "concentration" is one droplet of the colloid from the original LUDOX bottle (as delivered with you system) diluted 100x. If the signal is too strong, dilute further.
 +
 +Note that recording the IRF via scattering requires tuning the emission monochromator to the excitation wavelength. In filter based machines, e.g. FluoTime100 this means removing the emission bandpass or longpass filter. In microscopes, this is equivalent to replacing the detection filter with an OD filter (OD = optical density, grey filter) and recording the back scattering from e.g. a clean glass cover slip. Beware, on some confocal LSMs it is simply not possible to measure at the excitation wl. due to the restrictions introduced into the system software by the manufacturer. In case of measurements of solid samples in FT100 or FT200 spectrometers, the surface scattering is usually so strong that it is not necessary to mount a special sample for IRF. Tuning the mono to the exc. wl. and attenuating the excitation beam is sufficient. 
 +
 +==== Selected literature: ====
 +
 +Luchowski R., Kapusta P., Szabelski M., Sarkar P., Borejdo J., Gryczynski Z., Gryczynski I.\\ 
 +Förster resonance energy transfer (FRET)-based picosecond lifetime reference for instrument response evaluation \\ 
 +Measurement Science and Technology, Vol.20, 095601 (2009)\\ 
 +http://dx.doi.org/10.1088/0957-0233/20/9/095601
 +
 +
 +Luchowski R., Gryczynski Z., Sarkar P., Borejdo J., Szabelski M., Kapusta P., Gryczynski I.\\ 
 +Instrument response standard in time-resolved fluorescence\\ 
 +Review of Scientific Instruments, Vol.80, 033109 (2009)\\ 
 +http://dx.doi.org/10.1063/1.3095677
 +
 +
 +Szabelski M., Luchowski R., Gryczynski Z., Kapusta P., Ortmann U., Gryczynski I.\\ 
 +Evaluation of instrument response functions for lifetime imaging detectors using quenched Rose Bengal solutions\\ 
 +Chemical Physics Letters, Vol.471, p.153-159 (2009)\\ 
 +http://dx.doi.org/10.1016/j.cplett.2009.02.001
 +
 +
 +Szabelski M., Iliev D., Sarkar P., Luchowski R., Gryczynski Z., Kapusta P., Erdmann R., Gryczinski I.\\ 
 +Collisional quenching of Erythrosine B as a potential reference dye for impulse response function evaluation\\ 
 +Applied Spectroscopy, Vol.63, p.0363-0368 (2009)\\ 
 +http://www.ingentaconnect.com/content/sas/sas/2009/00000063/00000003/art00017
 +
 +
 +Szabelski M., Gryczynski Z., Gryczynski I.\\ 
 +Photophysical properties of novel fluorescein derivative and its applications for time-resolved fluorescence spectroscopy\\ 
 +Chemical Physics Letters, Vol.493, p.399-403 (2010)\\ 
 +http://dx.doi.org/10.1016/j.cplett.2010.05.061
 +
 +
 +The dye **pinacyanol iodide** has a very fast decay time (in methanol 6 ps) and very broad emission spectrum (starting at around 550nm), thus suitable for determination of the IRF in a very broad range.
 +
 +Pinacyanol iodide is also known as Quinaldine blue, 1,1' -Diethyl-2,2' -carbocyanine iodide, CAS Number:605-91-4. Pinacyanol bromide, CAS Number:2670-67-9 or Pinacyanol chloride CAS Number:2768-90-3 should work too, however, the dye must be pure, otherwise you may end up with measuring (long lifetime) impurity emission. For details see:
 +
 +van Oort B., Amunts A., Borst J. W., van Hoek A., Nelson N., van Amerongen H., Croce R.\\ 
 +Picosecond fluorescence of intact and dissolved PSI-LHCI crystals\\ 
 +Biophysical Journal, Vol.95, p.5851-5861 (2008)\\ 
 +http://dx.doi.org/10.1529/biophysj.108.140467
 +
 +
 +The dye **Allura Red** has also a remarkably short fluorescence lifetime of about 10 ps. This short lifetime does not depend on the emission (observation) wavelength (from 550 nm to 750 nm) and it is another candidate for impulse response measurements.
 +
 +Allura Red is Disodium 6-hydroxy-5-[(2-methoxy-5-methyl-4-sulfophenyl)azo]-2-naphthalenesulfonate, CAS Number: 25956-17-6, Colour Index Number 16035, EC Number 247-368-0, MDL number MFCD00059526, PubChem Substance ID: 24869338\\
 +
 +Rahul Chib, Sunil Shah, Zygmunt Gryczynski, Rafal Fudala, Julian Borejdo, Bogumil Zelent, Maria G Corradini, Richard D Ludescher, Ignacy Gryczynski\\ 
 +Standard reference for instrument response function in fluorescence lifetime measurements in visible and near infrared\\ 
 +Measurement Science and Technology, Vol.27, 027001 (2016)\\ 
 +http://dx.doi.org/10.1088/0957-0233/27/2/027001
 +
 +Pyridinium Styryl Dyes have also been used as IRF standard:
 +
 +Lei Li, Mengfang Chang, Hua Yi, Menghui Jia, Siaodan Cao, Zhongneng Zhou, Sanjun Zhang, Haifeng Pan, Chun-Wei Shih, Ralph Jimenez, Jianhua Xu\\
 +Using Pyridinium Styryl Dyes as the Standards of Time-Resolved Instrument Response\\
 +Applied Spectroscopy, Vol. 70(7), 1195-1201 (2016)\\
 +http://asp.sagepub.com/content/70/7/1195
 +
 +Even dye with lifetimes that are usually in the ns range can be used, if the lifetime can be quenched to a few ps. Quenching can be achieved using a saturated iodide solution. For details see:
 +
 +Mengwei Liu, Mengui Jia, Haifeng Pan, Lei Li, Mengfang Chang, Hua Ren, Françoise Argoul, Sanjun Zhang, Jianhua Xu\\
 +Instrument Response Standard in Time-Resolved Fluorescence Spectroscopy at Visible Wavelength: Quenched Fluorescein Sodium\\
 +Applied Spectroscopy, Vol. 68, p. 577-583 (2014)
 +http://dx.doi.org/10.1366/13-07236
 + 
 +
 +
 +
howto/how_to_measure_the_instrument_response_function_irf.txt · Last modified: 2023/09/07 22:55 by peter