howto:how_to_measure_the_instrument_response_function_irf
Differences
This shows you the differences between two versions of the page.
Both sides previous revisionPrevious revision | Next revisionBoth sides next revision | ||
howto:how_to_measure_the_instrument_response_function_irf [2015/06/03 14:58] – admin | howto:how_to_measure_the_instrument_response_function_irf [2016/01/04 13:46] – New dye, Allura Red added peter | ||
---|---|---|---|
Line 67: | Line 67: | ||
Biophysical Journal, Vol.95, p.5851-5861 (2008)\\ | Biophysical Journal, Vol.95, p.5851-5861 (2008)\\ | ||
http:// | http:// | ||
+ | |||
+ | |||
+ | The dye **Allura Red** has also a remarkably short fluorescence lifetime of about 10 ps. This short lifetime does not depend on the emission (observation) wavelength (from 550 nm to 750 nm) and it is another candidate for impulse response measurements. | ||
+ | |||
+ | Allura Red is Disodium 6-hydroxy-5-[(2-methoxy-5-methyl-4-sulfophenyl)azo]-2-naphthalenesulfonate, | ||
+ | |||
+ | Rahul Chib, Sunil Shah, Zygmunt Gryczynski, Rafal Fudala, Julian Borejdo, Bogumil Zelent, Maria G Corradini, Richard D Ludescher, Ignacy Gryczynski\\ | ||
+ | Standard reference for instrument response function in fluorescence lifetime measurements in visible and near infrared\\ | ||
+ | Measurement Science and Technology, Vol.27, 027001 (2016)\\ | ||
+ | http:// | ||
+ | |||
howto/how_to_measure_the_instrument_response_function_irf.txt · Last modified: 2023/09/07 22:55 by peter