User Tools

Site Tools


howto:how_to_measure_the_instrument_response_function_irf

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
Next revisionBoth sides next revision
howto:how_to_measure_the_instrument_response_function_irf [2019/11/04 09:45] – [Appropriate Count Rate for Measuring an IRF] buschmannhowto:how_to_measure_the_instrument_response_function_irf [2022/01/24 10:22] – [IRF measurement with KI and Erythrosine B] rezvani
Line 43: Line 43:
  prepare 1mL of saturated water solution of KI (potassium iodide)   prepare 1mL of saturated water solution of KI (potassium iodide) 
  add 0.17 mL of saturated water solution of Erythrosine B (at least 95% of purity)  add 0.17 mL of saturated water solution of Erythrosine B (at least 95% of purity)
- add 0.03 mL of 0.004 M KOH (potassium hydroxide) solution in order to achieve pH10  
 </code> </code>
  
Line 123: Line 122:
 Evaluation of instrument response functions for lifetime imaging detectors using quenched Rose Bengal solutions\\  Evaluation of instrument response functions for lifetime imaging detectors using quenched Rose Bengal solutions\\ 
 Chemical Physics Letters, Vol.471, p.153-159 (2009)\\  Chemical Physics Letters, Vol.471, p.153-159 (2009)\\ 
-http://dx.doi.org/10.1016/j.cplett.2009.02.001+https://www.sciencedirect.com/science/article/abs/pii/S0009261409001389
  
  
Line 129: Line 128:
 Collisional quenching of Erythrosine B as a potential reference dye for impulse response function evaluation\\  Collisional quenching of Erythrosine B as a potential reference dye for impulse response function evaluation\\ 
 Applied Spectroscopy, Vol.63, p.0363-0368 (2009)\\  Applied Spectroscopy, Vol.63, p.0363-0368 (2009)\\ 
-http://www.ingentaconnect.com/content/sas/sas/2009/00000063/00000003/art00017+https://www.osapublishing.org/as/viewmedia.cfm?uri=as-63-3-363&seq=0
  
  
Line 135: Line 134:
 Photophysical properties of novel fluorescein derivative and its applications for time-resolved fluorescence spectroscopy\\  Photophysical properties of novel fluorescein derivative and its applications for time-resolved fluorescence spectroscopy\\ 
 Chemical Physics Letters, Vol.493, p.399-403 (2010)\\  Chemical Physics Letters, Vol.493, p.399-403 (2010)\\ 
-http://dx.doi.org/10.1016/j.cplett.2010.05.061+https://www.sciencedirect.com/science/article/abs/pii/S0009261410007256
  
  
Line 145: Line 144:
 Picosecond fluorescence of intact and dissolved PSI-LHCI crystals\\  Picosecond fluorescence of intact and dissolved PSI-LHCI crystals\\ 
 Biophysical Journal, Vol.95, p.5851-5861 (2008)\\  Biophysical Journal, Vol.95, p.5851-5861 (2008)\\ 
-http://dx.doi.org/10.1529/biophysj.108.140467+https://www.sciencedirect.com/science/article/pii/S0006349508820012
  
  
howto/how_to_measure_the_instrument_response_function_irf.txt · Last modified: 2023/09/07 22:55 by peter