howto:calibrate_the_confocal_volume_for_fcs_using_the_fcs_calibration_script

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revisionBoth sides next revision
howto:calibrate_the_confocal_volume_for_fcs_using_the_fcs_calibration_script [2014/06/18 09:33] adminhowto:calibrate_the_confocal_volume_for_fcs_using_the_fcs_calibration_script [2015/11/03 14:18] admin
Line 19: Line 19:
 The diffusion coefficient of ATTO488 carboxilic acid has been measured to be 400 µm²/s at 25°C. Published diffusion coefficients of various fluorophores are summarized in the Technical Note "Absolute Diffusion Coefficients: Compilation of Reference Data for FCS Calibration" available from PicoQuant [see [[http://www.picoquant.com/images/uploads/page/files/7353/appnote_diffusioncoefficients.pdf]]]. This list is extensive, but does not claim to be complete.\\ The diffusion coefficient of ATTO488 carboxilic acid has been measured to be 400 µm²/s at 25°C. Published diffusion coefficients of various fluorophores are summarized in the Technical Note "Absolute Diffusion Coefficients: Compilation of Reference Data for FCS Calibration" available from PicoQuant [see [[http://www.picoquant.com/images/uploads/page/files/7353/appnote_diffusioncoefficients.pdf]]]. This list is extensive, but does not claim to be complete.\\
 Note that the diffusion coefficients of all dyes are temperature dependent.\\ Note that the diffusion coefficients of all dyes are temperature dependent.\\
-FCS measurements are usually point measurements acquired with very sensitive detectors ([[glossary:spad|SPAD]] or Hybrid PMT detectors).\\+FCS measurements are usually point measurements acquired with very sensitive detectors ([[glossary:spad|SPAD]] or [[glossary:Hybrid PMT]] detectors).\\
 FCS is a technique applicable over a wide range of concentrations, but ideal FCS concentrations are usually between 1 and 20 nM. FCS is a technique applicable over a wide range of concentrations, but ideal FCS concentrations are usually between 1 and 20 nM.
  