Line 22: Line 22:
  
  
-===== Measurement of Pile Up Limit =====+===== Measurement of the Pile Up Limit =====
  
  
Line 31: Line 31:
   * At a pixel dwell time of 10µs each pixel is displayed as one point in the time trace. At a pixel dwell time of 1 µs the average of 10 pixels are displayed as one point in the time trace.    * At a pixel dwell time of 10µs each pixel is displayed as one point in the time trace. At a pixel dwell time of 1 µs the average of 10 pixels are displayed as one point in the time trace. 
   * During the Test FLIM measurement the laser intensity is set that the time trace is only at rare instances hitting the red line.    * During the Test FLIM measurement the laser intensity is set that the time trace is only at rare instances hitting the red line. 
 +
 +===== Calculation of the count rate per pixel to determine the Pile-Up effect  =====
 +
 +For an already acquired FLIM measurement it is possible to determine the count rate per pixel to determine if a pile-up effect is present. 
 +Procedure:
 +  * Calculate the exposure time per pixel: Multiply the number of frames with the pixel dwell time
 +  * Determine the number of photons via the data reader (right mouse click) in a very bright pixel within the FastFLIM image
 +  * Calculate the count rate for this bright pixel: Divide the number of photons per pixel by the exposure time per pixel
howto/avoid_pile_up_effect_in_flim_measurements.txt · Last modified: 2015/06/03 15:11 by admin