Line 16: Line 16:
     *with [[Products:PicoHarp 300]] in [[glossary:T3-mode]]     *with [[Products:PicoHarp 300]] in [[glossary:T3-mode]]
         *not possible with SymPhoTime (32-bit version), because in SPT32 the Sync Divider is always active and set to 8         *not possible with SymPhoTime (32-bit version), because in SPT32 the Sync Divider is always active and set to 8
-        *instead, data acquisition is possible with native [[Products:PicoHarp 300]]/[[Products:HydraHarp 400]] softwares as well as with [[software:SymPhoTime64]], where the sync divider can be disabled. With a HydraHarp400, don't use use SYNC channel as input for one detector. Use e.g. input channels 1 and 2. SYNC signal must be disconnected. Unfortunately, analysis of HH T3-antibunching data is not implemented in SPT. +        
 +Instead, data acquisition is possible with native [[Products:PicoHarp 300]]/[[Products:HydraHarp 400]] softwares as well as with [[software:SymPhoTime64]], where the sync divider can be disabled. 
 + 
 +Using a PicoHarp300, connect one detector to Ch.0 (SYNC) and the other one to Ch.1 
 + 
 +With a HydraHarp400, don't use use SYNC channel as input for one detector. Use e.g. input channels 1 and 2. The SYNC should stay connected. 
 + 
 +Unfortunately, analysis of HydraHarp400 T3-mode antibunching data is not implemented in SPT. 
     *with [[Products:PicoHarp 300]]/[[Products:HydraHarp 400]] in [[glossary:T2-mode]]     *with [[Products:PicoHarp 300]]/[[Products:HydraHarp 400]] in [[glossary:T2-mode]]
-        *measurements are possible with the [[software:SymPhoTime64]]+        *measurements are possible with [[software:SymPhoTime64]]
  
 ==== Changing from Normal Acquisition (T3-mode) to Antibunching (T2-mode) ==== ==== Changing from Normal Acquisition (T3-mode) to Antibunching (T2-mode) ====
Line 24: Line 31:
 === Cabling: PicoHarp === === Cabling: PicoHarp ===
  
-     -Disconnect Sync from [[Products:PicoHarp 300]] Ch0+     -Disconnect Sync from [[Products:PicoHarp 300]] Ch.0
-     -Disconnect SPAD 1 Signal from Router ([[Products:PHR800]])+     -Disconnect SPAD 1 Signal from PHR800 Router
-     -Connect SPAD 1 to [[Products:PicoHarp 300]] Ch0 (Do not use the attenuator).+     -Connect SPAD 1 Signal to [[Products:PicoHarp 300]] Ch.0 only if it is NIM compatible. If the SPAD provides only TTL output pulses, you need to use the SIA400 adapter.
      -Adjustment of the CFD Level for Ch 0 (Sync Channel) might be necessary.       -Adjustment of the CFD Level for Ch 0 (Sync Channel) might be necessary. 
  