| - | ====== Determination of the Focal Width with the Focal Width Analysis ====== | ||
| - | |||
| - | |||
| - | ===== Summary ===== | ||
| - | |||
| - | |||
| - | This tutorial shows step-by-step, | ||
| - | |||
| - | ===== Step-by-Step Tutorial ===== | ||
| - | |||
| - | ==== Select a file and start the script ==== | ||
| - | |||
| - | * Start [[products: | ||
| - | * Open the " | ||
| - | |||
| - | **Note:** The " | ||
| - | |||
| - | **Response: | ||
| - | {{ determination_of_the_focal_width_with_the_focal_Image_1.png }} | ||
| - | |||
| - | * Highlight the file '' | ||
| - | |||
| - | **Note:** This file contains the image of a TetraSpeck Bead with 100 nm diameter. As this is smaller than the optical resolution, beads of this size can be used to measure the focal width in a microscope system. | ||
| - | |||
| - | {{ determination_of_the_focal_width_with_the_focal_Image_2.png }} | ||
| - | |||
| - | * Select the " | ||
| - | {{ determination_of_the_focal_width_with_the_focal_Image_3.png }} | ||
| - | |||
| - | **Note:** The drop down menu can be opened and closed by clicking on the grey button on the left side of the header of the drop down menu: {{grey_button.png}} | ||
| - | |||
| - | * Start the "Focal Width" script by clicking on " | ||
| - | {{ determination_of_the_focal_width_with_the_focal_Image_4.png }} | ||
| - | |||
| - | **Response: | ||
| - | {{ determination_of_the_focal_width_with_the_focal_Image_5.png? | ||
| - | |||
| - | **Note:** The window contains two different regions: | ||
| - | |||
| - | - **Left:** Image plotting analysis options. | ||
| - | - **Center/ | ||
| - | |||
| - | **Optional: | ||
| - | * Pixel binning by changing the " | ||
| - | * Select the second excitation peak by clicking on the white area of the "Set Time Gate" graph. | ||
| - | |||
| - | **Note:** The data in the file is a bead that was excited via pulsed interleaved excitation (PIE) using two perpendicularly polarized lasers, which were guided through a DIC prism. This causes a displacement of the beams, which is used in dual focus FCS (2fFCS) measurements.\\ | ||
| - | If data are acquired with a scanner controlled by the SymPhoTime 64 software (i.e. the piezo scanner in the MicroTime 200), | ||
| - | |||
| - | {{ determination_of_the_focal_width_with_the_focal_Image_6.png }} | ||
| - | |||
| - | * When finished, click " | ||
| - | * Click into the intensity center of the bead. | ||
| - | {{ determination_of_the_focal_width_with_the_focal_Image_7.png? | ||
| - | |||
| - | * The image fitting is already done and you can click on "Save Result" | ||
| - | {{ determination_of_the_focal_width_with_the_focal_Image_8.png }} | ||
| - | |||
| - | * Optional: Apply the same script to the file '' | ||
| - | |||